Description

Purpose of the biological safety cabinet class II type A2 (code 221.120):

  • Physical isolation (retention and controlled removal from the working area) of pathogenic biological agents (PBA) and microorganisms in order to prevent the possibility of infection by airborne droplets of personnel and contamination of the air of the working room and the environment;
  • Minimizing the risk of infection and cross-contamination of the product;
  • Ability to work with a small amount of toxic chemicals, radionuclides and remove odors of working agents with the obligatory connection of the box to an individual active exhaust ventilation system using an exhaust hood, which is supplied by an additional order.

Application of biological safety cabinet class II type A2:

The box is designed to equip individual workplaces of medical, pharmaceutical and other institutions that work with pathogenic biological agents and microorganisms in accordance with SP 1.3.2322-08, SP 1.3.2518-09, SP 1.3.3118-13.

Registration certificate of Roszdravnadzor No. FSR 2012/13259 dated 05.05.2012

Standard version of the microbiological safety cabinet BMB-II-“Laminar-S”-1.2

Working chamber :

front glass – hinged, glass material “triplex”, the mechanism for opening, closing and holding the glass in the open position is equipped with gas shock absorbers;

  • a damper for prevention of blow when closing front glass;

Tempered side windows

Inclined front surface of the box;

lighting of the working chamber;

  • 2 sockets in the working chamber of the box (on the right on the back wall);
  • air flow laminarizer made of fine-meshed polymer mesh;

a set of removable stainless steel worktops (three sections);

tabletop reinforcements (3 pcs.);

tabletop supports (2 pcs.);

Detachable stainless steel palm rest.

UV irradiation unit:

retractable (in non-working position it is located outside the working chamber, under its base);

  • metal protection of a lamp of UF-irradiation.
  • dampers to prevent shock when lowering the lifting gate of the working opening.

Cleaning system for incoming and outgoing air:

  • purification of the air entering the working chamber occurs through the supply HEPA filter H14;
  • purification of the air removed from the box occurs through the outlet HEPA filter H14;
  • For the movement of air flows inside the box, one fan is installed.

Elements of the control system:

  • electroautomatic system – microprocessor;

boxing control panel – push-button with LCD display;

  • electromagnetic reader and electromagnetic access keys;

air flow parameters sensors;

  • position sensors of front glass and block of UF-irradiation;

individual fuse for the power supply line of the main functions of the box (fan, lighting, UV irradiation);

individual fuse for sockets installed in the working chamber;

Detachable power cable with lock.

Items to adjust and check:

  • the balancing damper of an exhaust aperture for regulation of air streams;
  • Built-in air sampling fitting before HEPA filters to check their integrity.

Stand :

stand – frame with a shelf for legs;

transport wheels are provided for moving the box on the stand;

  • For stationary installation of the box at the place of operation, screw supports are provided.

Main characteristics of the biological safety LAMINAR BOX class II type A2
BMB-II-“Laminar-S”-1,2

The class of air purity in the working chamber in terms of the concentration of suspended particles (aerosols)
according to GOST ISO 14644-1-2002, not less than
5 ISO
Boxing class according to GOST R EN 12469-2010, NSF/ANSI 49 II
Box type according to NSF/ANSI 49 A2
Class of installed HEPA filters according to GOST R EN 1822-1-2010 H14
The average speed of the air flow entering the box through the working opening, m/s 0.47±0.03
The average speed of the downward air flow in the working chamber of the box, m/s 0.35+0.01
Illumination of the working area (integral value determined over the entire area of ​​the working area), Lux, not less than 1000
Degree of air recirculation in the box, % ≈70

Main parameters and dimensions of the biological safety LAMINAR BOX class II type A2
BMB-II-“Laminar-S”-1,2

Product number 1R-B.001-12.0
Overall dimensions of the box assembly with stand (WxDxH), mm 1200x770x2150
Height of the box assembled with a stand and with an installed umbrella*, mm 2460
Height from floor to work surface, mm 805
Dimensions of the working chamber of the laminar box (WxDxH), mm 1105x610x750
Weight of the box assembled with a stand (net), kg, no more 235
Power consumed by the box (excluding the load on the built-in sockets), W, not more than 180/110**
Total maximum allowable load on built-in sockets, W, no more 1000
Sound pressure level at a distance of 1 m from the box, dB(A), no more 55***
Productivity for clean air supplied to the working chamber of the box, m3/h 795-817