Description
Purpose of the biological safety cabinet class II /type A2/
(code 221.180 ex. 220.180):
physical isolation (retention and controlled removal from the working area) of pathogenic biological agents (PBA) and microorganisms in order to prevent the possibility of infection by airborne droplets of personnel and contamination of the air of the working room and the environment;
minimizing the risk of infection and cross-contamination of the product;
Possibility to work with a small amount of toxic chemicals, radionuclides and remove odors of working agents with the obligatory connection of the box to an individual active exhaust ventilation system using an exhaust hood, which is supplied by an additional order;
- the box is designed to equip individual workplaces of medical, pharmaceutical and other institutions that work with pathogenic biological agents and microorganisms in accordance with SP 1.3.2322-08, SP 1.3.2518-09, SP 1.3.3118-13.
Registration certificate of Roszdravnadzor No. FSR 2012/13259 dated 05.05.2012
Standard version of the biological safety cabinet class II /type A2/
(code 221.180 ex. 220.180)
Working chamber
front glass – hinged, glass material “triplex”, the mechanism for opening, closing and holding the glass in the open position is equipped with gas shock absorbers;
dampers to prevent shock when closing the front glass;
Tempered side windows
Inclined front surface of the box;
lighting of the working chamber;
- 4 sockets in the working chamber of the box (2 sockets on the left and right on the back wall);
- air flow laminarizer made of fine-meshed polymer mesh;
a set of stainless steel worktops (three sections);
tabletop reinforcements (3 pcs.);
tabletop supports (2 pcs.);
Two removable stainless steel palm rests.
UV block
retractable (in non-working position it is located outside the working chamber, under its base);
- metal protection of a lamp of UF-irradiation;
- dampers to prevent shock when lowering the lifting gate of the working opening.
Cleaning system for incoming and outgoing air
- purification of the air entering the working chamber occurs through two supply HEPA filters H14;
- Purification of the air removed from the box occurs through the outlet HEPA filter H14.
Elements of the control system
- electroautomatic system – microprocessor;
boxing control panel – push-button with LCD display;
- electromagnetic reader and electromagnetic access keys;
air flow parameters sensors;
- position sensors of front glass and block of UF-irradiation;
individual fuse for the power supply line of the main functions of the box (lighting, fan, UV irradiation);
individual fuse for sockets installed in the working chamber;
- non-removable power cable (built into the base of the box).
Elements for adjustment and verification
- built-in air sampling fittings in front of HEPA filters to check their integrity.
Stand
stand – frame with a shelf for legs;
transport wheels are provided for moving the box on the stand;
- For stationary installation of the box at the place of operation, screw supports are provided.
The main characteristics of the biological safety LAMINAR BOX class II / type A2 /
BMB-II-“Laminar-S”-1,8 NEOTERIC
The class of air purity in the working chamber in terms of the concentration of suspended particles (aerosols) according to GOST ISO 14644-1-2002, not less than | 5 ISO |
Boxing class according to GOST R EN 12469-2010, NSF/ANSI 49 | II |
Box type according to NSF/ANSI 49 | A2 |
Class of installed HEPA filters according to GOST R EN 1822-1-2010 | H14 |
The average speed of the downward air flow in the working chamber of the box, m/s | 0.35+0.01 |
The average speed of the air flow entering the box through the working opening, m/s | 0.47±0.03 |
Illumination of the working area (integral value determined over the entire area of the working area), Lux, not less than | 1000 |
Degree of air recirculation in the box, % | ≈70 |
Main parameters and dimensions of the biological safety LAMINAR BOX class II / type A2 /
BMB-II-“Laminar-S”-1,8 NEOTERIC
Product number | 1R-B.001-18.0 |
Overall dimensions of the box assembly with stand (WxDxH), mm | 1800x770x2092 |
Height of the box assembled with a stand and with an installed umbrella*, mm | 2402 |
Height from floor to work surface, mm | 805 |
Dimensions of the working chamber (WxDxH), mm | 1705x610x750 |
Power consumed by the box (excluding the load on the built-in sockets), W, not more than | 590/150** |
Total maximum allowable load on built-in sockets, W, no more | 1000 |
Productivity for clean air supplied to the working chamber of the box, m3/h | 1210-1245 |
Sound pressure level at a distance of 1 m from the box, dB, no more | 59 |