Description

Defibrillator-monitor DKI-N-10 “Axion”

With a thermal printer and a combined power supply, it is used for electropulse therapy for cardiac arrhythmias. The defibrillator-monitor DKI-N-10 “Axion” is used in medical hospitals, cardiological dispensaries, to equip ambulance and emergency medical teams.

  • The defibrillator consists of two parts: a wearable part – weight 6 kg and a charger – weight 2 kg
  • Defibrillation pulse – bipolar trapezoidal with a limited duration (BTOD) asymmetrical with the ratio of the amplitudes of the currents of the negative and positive half-waves (0.5 ± 0.1) at a load of 50 Ohm
  • Can be used for adults and children
  • Voice accompaniment of the operator’s actions and the operation of the device
  • Automatic reset of stored energy to internal load 30 seconds after set
  • Possibility of receiving and recording ECG
  • Built-in thermal printer – width at least 57 mm
  • The time of continuous operation of the device in the monitoring mode – at least 7 hours from replaceable batteries (2 pcs.), when operating from a 220 V network, from the car’s on-board network (12-30 V) – at least 168 hours
  • Ability to charge two batteries at the same time
  • Display of all information on the LCD before and after defibrillation
  • cardiosynchronized
  • Manual control.

Specifications

Dimensions: Length 380 mm, width 175 mm, height 270 mm
Screen dimensions, 115 × 86 mm
Energy build-up time 200 J, no more than 6 s
Energy build-up time 360 ​​J, no more than 10 s
Number of energy discharges 360 J from a fully charged battery, not less than 40
ECG channel sensitivity, 5, 10, 20 mmV
Image movement speed, 12.5, 25 mms
Pulse energy, 5, 10, 25, 50, 75, 100, 150, 200, 250, 300, 360 J
Power consumption , not more than 250 W
Electrical safety class II, type BF