Gas preparation unit BPG-5M
The BPG-5M gas preparation unit is designed for preliminary preparation (filtration, drying, stabilization of operating parameters) of gases before they are supplied mainly to devices for flame spectral analysis and other devices that require purification of working gases.
The main area of application is the preparation of gases for flame atomic absorption spectrometers, including KVANT-2m , KVANT-2AT , KVANT -2A , KVANT.Z , KVANT.Z1 , mercury-hydride generators (GRG) and other analytical equipment.
The unit is designed for the preparation of the following gases:
propane,
acetylene,
nitrous oxide,
argon,
compressed air.
Specifications
Absolute fineness of filtration, µm | 5 |
Mean time between failures, hour, not less | 2500 |
Full service life, years, not less | 6 |
Overall dimensions of the block, mm, no more | 510 x 575 x 130 |
Block mass, kg, no more | 5 |
Gas inlet pressure range:
Gas |
Inlet pressure, MPa |
Inlet pressure, atm |
Propane |
from 0.04 to 0.12 |
from 0.4 to 1.2 |
Acetylene |
from 0.14 to 0.18 |
from 1.4 to 1.8 |
Nitrous oxide |
from 0.50 to 0.60 |
from 5.0 to 6.0 |
Air |
from 0.40 to 0.60 |
from 4.0 to 6.0 |
Argon |
from 0.30 to 0.45 |
from 3.0 to 4.5 |
Block modules provide the following gas flow rates :
Spectrometer air, l/min, not less than | 20 |
Air for GRG, l/min, not less than | 5 |
Nitrous oxide, l/min, not less than | 20 |
Acetylene, l/min, not less than | 7 |
Propane, l/min, not less | 1 |
Argon, l/min, not less | 5 |
Operating conditions
Ambient temperature, °С | from +10 to +35 |
Maximum relative humidity at +25 °С, % | 80 |
Atmospheric pressure, MPa (atm) | from 0.084 to 0.107 (from 0.84 to 1.07) |